An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Oct 20, 2022 Infusion Pumps Nationwide View Details β†’

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Oct 17, 2022 Infusion Pumps Nationwide View Details β†’

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

Oct 13, 2022 Infusion Pumps Nationwide View Details β†’

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Sep 20, 2022 Infusion Pumps Nationwide View Details β†’

Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.

Sep 20, 2022 Infusion Pumps Nationwide View Details β†’

Dimension VistaΒΏ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.

Aug 24, 2022 Infusion Pumps Nationwide View Details β†’

Mislabeling: the corrugate label, lidstock, banner card, and peel sticker accompanying the affected product incorrectly displays a double lumen 5.5 French catheter (and related priming volumes and kit contents for product code ASK-41552-JHVA1), instead of the packaged single-lumen 4.5 French catheter (and related priming volumes and kit contents for product code ASK-41541-JHVA1).

Sep 14, 2022 Infusion Pumps Nationwide View Details β†’