It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl
Infusion Pumps
🏥 Medical Devices • 4,262 recalls
Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will
ACIST Medical Systems Inc. The ACIST Angiographic Injection System is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.
Acist Medical Systems
ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w
During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
ADVIA Centaur Vitamin D Calibrator 2-pack; Reference Number:10493589. For use in the quantitative determination of Vit D.
Siemens Healthcare Diagnostics
Negative bias in patient samples for Vitamin D
ADVIA Centaur Vitamin D Calibrator 6-pack; Reference Number:10630911. For use in the quantitative determination of Vit D.
Siemens Healthcare Diagnostics
Negative bias in patient samples for Vitamin D
Kerr Corporation is initiating this recall because raw material used to produce the Maxcem Elite is causing the material to prematurely polymerize in the syringe. The product could be difficult to extrude or demonstrate a shorter work-time as the material may prematurely set prior to placing restorations on the tooth or core build-up. As a result, the restoration may not be seated properly.
Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the pump's internal electronics. This moisture damage can prevent the pump's buttons from working properly or can cause the pump to alarm.
LHB, First Aid Type IV Belt Type, NSN #6545-01-010-7754. Kit number: 114016. First Aid Kit.
Lighthouse For The Blind
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility claim is only applicable to the applicator. The antiseptic solution that is applied to the skin is a cutaneous disinfectant, and is not sterili
GE Healthcare has become aware of an issue associated with the base casting for the articulated arm of your VMX and VMX Plus Mobile X-ray systems that may impact Operator or Patient safety.
G7 HPLC. Used as an aid in the detection and presumptive identification of abnormal hemoglobins in whole blood.
Tosoh Bioscience
The Beta-thalessemia Mode of the G7 Analyzer-Beta Thalassemia Mode is an automated High Performance Liquid Chromatography (HPLC) system that separates and reports HbF and HbA2 quantitative percentages in whole blood. The firm received complaints that the A2 peak could not be detected with the G7 Beta-thalessemia columns.
Synthes is initiating a voluntary recall of the Flexible Grip (Part number 355.28) which is part of the Universal Nail System, due to the potential lack of the ability to thoroughly clean the instrument.
LHB, Perishable Items Pack Type IV. Kit number: 2114016-100. First Aid Kit.
Lighthouse For The Blind
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility claim is only applicable to the applicator. The antiseptic solution that is applied to the skin is a cutaneous disinfectant, and is not sterili
Single Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).
Siemens Healthcare Diagnostics
Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.
LHB, First Aid Type I Individual Kit, NSN: 6545-00-656-1092, antiseptics, Kit number: 2114002. First Aid Kit.
Lighthouse For The Blind
The product contains an iodine 10% Ampule (10% Povidone Iodine swab) which is being recalled by the firm's distributor. The labeling of the iodine 10% Ampule labels state that the contents of the package are sterile and does not clarify that this sterility claim is only applicable to the applicator. The antiseptic solution that is applied to the skin is a cutaneous disinfectant, and is not sterili
ELEKTA MOSAIQ SW version below 2.41 Product Usage: The intended use of the SIEMENS branded ARTISTE and ONCOR family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Siemens Medical Solutions USA
A potential safety risk with the MOSAIQ Oncology Information System that has been identified by the manufacturer IMPAC Medical Systems, Inc.
Horiba Medical is recalling ABX Micros IM2 Data Management System because it may display incorrect patient test results under unanticipated operating conditions.
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.