CS 100i Intra-Aortic Balloon Pump
Maquet Datascope Corp - Cardiac Assist Division
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.