DePuy Orthopaedics, Inc. is issuing a voluntary recall of ten lots of the PinnacleΒΏ Cancellous Screw due to an incorrect description on the label. The label description reads Pinnacle Cannulated Screw and should read Pinnacle Cancellous Screw.

Nov 28, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

Advantage Workstation Volume Share 2 (versionAW4.4_04) with Volume Viewer 3 and 3.1 (version7.0 to version7.5 and version 8.1) and only when or more of these applications is installed: " VesselIQ Xpress or AVA Xpress " CardIQ Xpress Pro or Plus " CardEP " CardIQ Fusion PET or SPECT Advantage Workstations Volume Share (v.AW4.3) and AW4.2 (version4.2-05 and above) provided with Volume Viewer 2(version6.0 to version6.11) and only when one or both of these applications is installed: " CardIQ Xpress Pro or Plus " CardEP Advantage Workstations AW Volume Share 2 (AW version 4.4). ONLY IF ONE OF FOLLOWING APPLICATIONS IS INSTALLED: ΒΏ VesselIQ Xpress or AVA Xpress ΒΏ CardIQ Xpress Pro or Plus ΒΏ CardEP ΒΏ CardIQ Fusion PET or SPECT provided with Volume Viewer 3 and 3.1 from versions 7.0 to 7.5 and 8.1. To verify the version installed on your system: Select the Admin menu from the Patient list page. Then select Display Configuration. You will find the version of your system in Installed Application(s). This version will appear as vxtl_7.x or vxtl_8.x. Intended use: K993792: Smart Vessel Analysis (Smart VA) is a software post-processing option for the Advantage Workstation (AW) platform, which can be used in the analysis of 3D angiography data. It provides a number of display, measurement and batch filming/archive features and will aid physicians in studying user-selected vessels for stenosis analysis, pre/post stent planning and directional vessel tortuosity visualization. K031261: Card EP is a post processing software option for the Advantage Workstation (AW) Platform. This product can be used for the analysis of CT angiographic images for the assessment of the heart to include the atria, pulmonary veins, and coronary sinus. It provides quantitative analysis tools which include a number of display, measurement and model export capabilities. This product can be used to aid trained physicians in the visualization and assessment of cardiac anatomy. K041267: CardIQ Analysis III is a CT image analysis software package, which allows the visualization of 2D and 3D medical image data of the heart derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular disease assessment. It provides functionally for 2D/3D rendering, assessment of calcified and non-calcified plaque to determine the densities of the plaque within a coronary artery, ventricular function of the heart, and measurement tools to detect coronary artery stenosis. This product can be used to aid a trained physician to process, render, review, archive, print and visualizing cardiac anatomy and coronary vessels. CardIQ Analysis II will run on the AW workstation, scanner operator console and PACS system. K061370: CardIQ Fusion is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology, aid in the assessment of functional data e.g. PET perfusion, and aid in tailoring treatment plans based on the fused anatomical and functional information. Anatomical data could be from a set of Computed Tomography (CT) Angiographic images while functional data could be from PET, SPECT, or processed CT data. CardIQ Fusion is a software post-processing package for the Advantage Workstation (AW) platform, PET/CT and CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic cardiac images/data providing a number of display, measurements and batch filming/archive features to study user-selected vessels. Also included is the capability to visualize reformatted CT/PET/SPECT perfusion and viability data. Finally, it provides different ways of visualizing CT anatomy fused with PET/SPECT functional information. With CardIQ Fusion, clinicians have the opportunity to overlay functional information over the (CT) anatomy of a patient's heart, and thus, they can potentially tailor their decision for that particular patient. CardIQ Fusion provides the visualization of the vessels in severa

GE Healthcare

Class I - Dangerous

GE Healthcare has became aware of two issues with the Advantage Workstation. 1) A possible mismatch between the label of the tracked vessel and the underlying image associated with the cardio- vascular applications of the Advantage workstation that may impact patient safety. 2) A refresh problem of the stenosis / aneurysm measurement tools associated with the cardio-vascular applications of

May 29, 2012 Implants & Prosthetics Nationwide View Details β†’

The Back Table Cover component in the custom pack was placed in the non-sterile portion of the pack when it should have been in the sterile portion.

Sep 20, 2012 Implants & Prosthetics Nationwide View Details β†’

Titan Spine LLC, is conducting a recall on the Endoskeleton TA VBR products. The labeling accompanying this device contains incorrect indications for use.

Oct 19, 2012 Implants & Prosthetics Nationwide View Details β†’

Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.

Nov 15, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’

All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening

Nov 14, 2012 Implants & Prosthetics Nationwide View Details β†’

An Engineering Analysis has determined the Solution Calcar Stem has the potential to result in impingement with a variety of heads. This impingement may not allow the stem and head to lock as intended. The amount of possible impingement varies depending on the femoral head. Worst case condition: Calcar stem with -2mm offset articuleze maximum of 0.937mm radial impingement. It is evident that a

Dec 3, 2012 Implants & Prosthetics Nationwide View Details β†’