This recall has been identified due to the batch being mislabeled, with the packaged component's size not corresponding to the label. Therefore, the product intended to be used for the procedure may not be included in the packaging resulting in surgical delay as replacement product is retrieved.
Implants & Prosthetics
🏥 Medical Devices • 4,540 recalls
There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.
An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.
The affected lots are being recalled because the epiphysis may not assemble to the stem, which may cause a surgical delay.
A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.
The affected lots are being recalled because the epiphysis may not assemble to the stem, which may cause a surgical delay.
OTTO BOCK Kenevo knee joints Model 3C60=ST
Otto Bock Healthcare Product
Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
OTTO BOCK Kenevo knee joints Model 3C60
Otto Bock Healthcare Product
Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Ziptight Ankle Syndesmosis Fixation Device, Titanium Implant Titanium Implant for surgical repairs of soft tissue.
Zimmer Biomet
Ziptight Ankle Syndesmosis Fixation was packaged without the tip protectors.
The 2nd Assist Knee Positioner
Zimmer Biomet
Zimmer Biomet is conducting a medical device recall for the 2nd Assistant Knee Positioner due to the potential for the sterile packaging barrier to be breached during distribution.
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct expiration date should be 2022-06.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
Knee implant components (tibial bearings) are labeled with the incorrect size.