This recall has been identified due to the batch being mislabeled, with the packaged component's size not corresponding to the label. Therefore, the product intended to be used for the procedure may not be included in the packaging resulting in surgical delay as replacement product is retrieved.

Feb 2, 2018 Implants & Prosthetics Nationwide View Details →

There is an increase in the failure rate of certain Anode Drive Units (ADU5) used in these products. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may also generate a peculiar burning smell that may be noticed. In some instances this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality is reduced.

Jan 4, 2018 Implants & Prosthetics Nationwide View Details →

When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.

Sep 8, 2017 Implants & Prosthetics View Details →

731 Series Ventilators running software version 05.20.00 The devices in the ZOLL ventilator are indicated for use in the management of infant through adult patients weighing ;:: 5 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. They are appropriate for use in hospitals, outside the hospital, during transport and in severe environments where they may be exposed to rain, dust, rough handling, and extremes in temperature and humidity. With an appropriate third-party filter in place, they may be operated in environments where chemical and/or biological toxins are present. When marked with an MRI conditional label, they are suitable for use in an MRI environment with appropriate precautions. The ventilators are intended for use by skilled care providers with knowledge of mechanical ventilation, emergency medical services (EMS) personnel with a basic knowledge of mechanical ventilation, and by first responders under the direction of skilled medical care providers. The EMV+¿ and Eagle II (with and without MRI label) have a full range of ventilation modes (AC, SIMV with or without PS, CPAP with or without PS with PPC or NPPV-PPV). The AEV¿ provides specific modes consistent with pre-hospital care provider s operating procedures (AC, CPAP with PS and NPPV-PPV). The ZOLL ventilators are a small, extremely durable, full-featured portable mechanical ventilator designed to operate in hospitals or severe and under-resourced environments. It can be used in prehospital (ALS, ATLS, ACLS), field hospitals, and hospital settings. Easy¿ to-use, durable,

ZOLL Medical

Class I - Dangerous

A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.

Jun 30, 2017 Implants & Prosthetics View Details →
Class I - Dangerous

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

Sep 12, 2017 Implants & Prosthetics Nationwide View Details →

OTTO BOCK Kenevo knee joints Model 3C60

Otto Bock Healthcare Product

Class I - Dangerous

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

Sep 12, 2017 Implants & Prosthetics Nationwide View Details →

Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.

Dec 20, 2017 Implants & Prosthetics Nationwide View Details →

The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.

Aug 1, 2017 Implants & Prosthetics Nationwide View Details →
Class I - Dangerous

Zimmer Biomet is conducting a medical device recall for the 2nd Assistant Knee Positioner due to the potential for the sterile packaging barrier to be breached during distribution.

Sep 7, 2017 Implants & Prosthetics View Details →

Versys 6 inch Beaded Full Coat Plus Hip Prosthesis. Labeled as follows: a. VERSYS 6 INCH BEADED FC 19X160MM STD BODY STD NECK (Item No. 00894301906); b. VERSYS 6 INCH BEADED FC 21X160MM STD BODY STD NECK (Item No. 00784302106); c. VERSYS 6 INCH BEADED FC 22X160MM STD BODY STD NECK (Item No. 00784302206); d. VERSYS 6 INCH BEADED FC STEM 19X160MM LM (Item No. 00784301936); e. VERSYS 6 INCH BEADED FC STEM 20X160MM LM (Item No. 00784302036); f. VERSYS 6 INCH BEADED FC STEM 21X160MM LM (Item No. 00784302136); g. VERSYS 6 INCH BEADED FC STEM 22X160MM LM (Item No. 00784302236); h. VERSYS 6 INCH BEADED FC 14X160MM STD BODY STD NECK (Item No. 00784301406); i. VERSYS 6 INCH BEADED FC 15X160MM STD BODY STD NECK (Item No. 00784301506); j. VERSYS 6 INCH BEADED FC 16X160MM STD BODY STD NECK (Item No. 00784301606); k. VERSYS 6 INCH BEADED FC 17X160MM STD BODY STD NECK (Item No. 00784301706); l. VERSYS 6 INCH BEADED FC 18X160MM STD BODY STD NECK (Item No. 00784301806); m. VERSYS 6 INCH BEADED FC 18X160MM STD BODY EXT NECK (Item No. 00784301826); n. VERSYS 6 INCH BEADED FC 19X160MM STD BODY EXT NECK (Item No. 00784301926); o. VERSYS 6 INCH BEADED FC 20X160MM STD BODY STD NECK (Item No. 00784302006); p. VERSYS 6 INCH BEADED FC 20X160MM STD BODY EXT NECK (Item No. 00784302026); q. VERSYS 6 INCH BEADED FC 21X160MM STD BODY EXT NECK (Item No. 00784302126); r. VERSYS 6 INCH BEADED FC 22X160MM STD BODY EXT NECK (Item No. 00784302226); s. VERSYS 6 INCH BEADED FC 17X160MM LM BODY XEXT NECK (Item No. 00784301746); t. VERSYS 6 INCH BEADED FC 17X160MM LM BODY EXT NECK (Item No. 00784301756 u. VERSYS 6 INCH BEADED FC STEM 18X160MM LM (Item No. 00784301836); v. VERSYS 6 INCH BEADED FC 18X160MM LM BODY XEXT NECK (Item No. 00784301846); w. VERSYS 6 INCH BEADED FC 18X160MM LM BODY EXT NECK (Item No. 00784301856); x. VERSYS 6 INCH BEADED FC 19X160MM LM BODY XEXT NECK (Item No. 00784301946); y. VERSYS 6 INCH BEADED FC 19X160MM LM BODY EXT NECK (Item No. 00784301956); z. VERSYS 6 INCH BEADED FC 20X160MM LM BODY XEXT NECK (Item No. 00784302046); aa. VERSYS 6 INCH BEADED FC 20X160MM LM BODY EXT NECK (Item No. 00784302056); ab. VERSYS 6 INCH BEADED FC 21X160MM LM BODY XEXT NECK (Item No. 00784302146); ac. VERSYS 6 INCH BEADED FC 21X160MM LM BODY EXT NECK (Item No. 00784302156); ad. VERSYS 6 INCH BEADED FC 22X160MM LM BODY XEXT NECK (Item No. 00784302246); ae. VERSYS 6 INCH BEADED FC 22X160MM LM BODY EXT NECK (Item No. 00784302256) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.

Zimmer Biomet

Class I - Dangerous

The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.

Aug 1, 2017 Implants & Prosthetics Nationwide View Details →

The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct expiration date should be 2022-06.

Aug 25, 2017 Implants & Prosthetics View Details →