There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.
Diagnostic Equipment
π₯ Medical Devices β’ 6,554 recalls
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENE SP devices.
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
FDA has informed Micro-Tech that the Elastic Traction Device is not appropriately classified as a Class I device and requires 510(k) clearance to be marketed in the US.
v|tome|x L Series non-destructive testing cabinet x-ray systems
Ge Inspection Technologies
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROSΒΏ Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROSΒΏ Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control
Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001
Siemens Healthcare Diagnostics
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation occurs.
Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000
Siemens Healthcare Diagnostics
Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
Potential for cleaning solution intrusion into endoscope
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.
Stryker Zyphr Disposable Cranial Perforator, Large 14/11 mm- is a sterile, single use cutting accessory intended for cutting an 11 mm diameter access hole through the cranium of adult patients. Catalog Number: 5100-060-001
Stryker Instruments Div. of Stryker
Potential for the inner bit to contain a crack, that may or may not be visible, which may lead to metal fragments entering the surgical site and/or delayed disengagment during use.
Strain should be VIM-1 positive but is confirmed to be VIM-1 negative.
Flow sensors could have damaged tubes in the form of small punctures or cuts.
There is a software problem which affects the DSA Roadmap application on ARTIS Icono and ARTIS Pheno systems with software version VE20B. In rare cases, when the DSA acquisition has been started while the C-Arm or table is moving, there may be a shift between the image acquired at the start position and the image (i.e. DSA vessel map) taken at the reference position. This may lead to a DSA vessel map being overlaid to a subtracted fluoro image at a position that is not accurate.. However, if the displacement is slight or at an unfavorable plane, the user might rely on incorrect visualization of the catheter relative to the vessel map. This could imply danger to the patient.
Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device).