Diagnostic Equipment

🏥 Medical Devices • 6,627 recalls

Advanced Sterilization Products (ASP) is recalling certain lots of the STERRAD CYCLESURE 24 Biological Indicator (BI) product because it does not have adequate data to support the entire duration of the product's labeled shelf-life.

Jun 28, 2012 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and

Mar 27, 2007 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 8800 Flexview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mod

Nov 20, 2006 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is defective and subject to early life failure.

Feb 22, 2008 Diagnostic Equipment View Details →

One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.

Jun 8, 2012 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo

Nov 20, 2006 Diagnostic Equipment View Details →

Sigma is a safety alert for the Master Drug Library (MDL) Editor software associated with the SIGMA Spectrum Infusion Pump System because it may result in a facility drug library that does not represent common clinical practices.

Apr 23, 2012 Diagnostic Equipment Nationwide View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Nov 19, 2007 Diagnostic Equipment View Details →

A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.

May 14, 2012 Diagnostic Equipment Nationwide View Details →

Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the test card lot (approximately 3000 cards) could potentially report low glucose results.

Jun 28, 2011 Diagnostic Equipment Nationwide View Details →

There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.

Jun 12, 2012 Diagnostic Equipment View Details →

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Nov 19, 2007 Diagnostic Equipment View Details →

The recall was initiated because Critical Diagnostics has confirmed that it is possible to obtain inaccurate test results for specimens when using the Presage ST2 Assay kits.

Feb 28, 2012 Diagnostic Equipment View Details →

In the Abbott Molecular Bar Code Scanner User's Guide, List Number 6L88-02/09, the section Modifying the Keyboard Format for Bar Code Scanner LN 6L89-01 contains incorrect bar codes that can prevent the correct configuration of the bar code scanner language format. The Abbott m2000 System is intended for use in performing nucleic acid testing in clinical laboratories. It is comprised of the Ab

Jan 31, 2012 Diagnostic Equipment View Details →

Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.

Jun 22, 2012 Diagnostic Equipment Nationwide View Details →

Defibrillator/monitor cannot analyze an ECG report during acquisition of 12 lead ECG

Jun 20, 2012 Diagnostic Equipment View Details →

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details →

The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.

May 23, 2012 Diagnostic Equipment Nationwide View Details →

Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.

Sep 8, 2011 Diagnostic Equipment Nationwide View Details →