πŸ”¬

Diagnostic Equipment

πŸ₯ Medical Devices β€’ 6,627 recalls

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Feb 18, 2013 Diagnostic Equipment Nationwide View Details β†’

Vital Images has found a potential error in measurement generated through the software which affects all versions of Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3. This error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/axia

Feb 5, 2013 Diagnostic Equipment View Details β†’

Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.

Nov 8, 2012 Diagnostic Equipment Nationwide View Details β†’

Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values

Nov 27, 2012 Diagnostic Equipment View Details β†’

Risk that air may be introduced into the device which may lead to an air embolism.

Feb 27, 2013 Diagnostic Equipment Nationwide View Details β†’

A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.

Jan 21, 2013 Diagnostic Equipment Nationwide View Details β†’

Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the on/off button. The overheating may result in discomfort ranging from too hot up to the potential for a burn.

Feb 27, 2013 Diagnostic Equipment Nationwide View Details β†’

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Dec 13, 2012 Diagnostic Equipment View Details β†’

Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.

Jan 28, 2013 Diagnostic Equipment View Details β†’

The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).

Dec 20, 2012 Diagnostic Equipment Nationwide View Details β†’

The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.

Dec 18, 2012 Diagnostic Equipment Nationwide View Details β†’

BD is conducting a recall of the BD Vacutainer Plus Plastic Plasma Separator Tube (PST) with Light Green/BD Homogard Closure due to a deformed top of the tube reservoir reported in two customer complaints.

Mar 5, 2012 Diagnostic Equipment Nationwide View Details β†’

Incorrect views and measurements possible. GE Healthcare became aware of multiple potential safety issues involving the Advantage workstation, Volume Viewer Plus and Volume Analysis.

May 25, 2012 Diagnostic Equipment Nationwide View Details β†’

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Oct 24, 2012 Diagnostic Equipment Nationwide View Details β†’