The firm is initiating this recall because the implied or stated indications in their labeling associated with Harmonie software have not yet been cleared by the FDA for sale in the USA.
Diagnostic Equipment
🏥 Medical Devices • 6,627 recalls
Glucose test strips reports no result message when tested
Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 51493 (UK Only)
Nova Biomedical
Glucose test strips reports no result message when tested
When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.
Transonic Systems Inc. have received complaints of blood leakage involving the AV Loop Kit.
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Roche Diagnostics Operations
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
ADVIA Centaur Vitamin D Assay (100 tests); Reference Number: 10491994. For use in the quantitative determination of Vit D.
Siemens Healthcare Diagnostics
Negative bias in patient samples for Vitamin D
Meter kit outer boxes containing 10 ct. test strips state incorrect storage temperature range.
The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).
Roche Diagnostics Operations
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.
The recall was initiated because Alere San Diego has an update for the Alere Cholestech LDX ALT AST cassette, and it has the potential for humidity changes to impact results for certain analytes on the Alere Cholestech LDX System.
Aplio 500/400/300 Diagnostic Ultrasound System (TUS-500/400/300). Device Listing Number: Dl53118. Diagnostic Ultrasound system.
Toshiba American Medical Systems
Toshiba Medical Systems has received reports of startup issues with the Toshiba TUS-500 and TUS-300 ultrasound systems. The startup issue is predominant when attempting to bring the system up from standby mode when the system is used remotely. It has been determined that this may potentially cause damage to the system.
The recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.
The recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic medical record (EMR) or the Alaris Auto-ID module for a continuous infusion with a bolus dose option may result in a loss of information from the Guardrails Data s
GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System.
GE Healthcare
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Siemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used as an in vitro diagnostic product.
Siemens Healthcare Diagnostics
Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL.
ADVIA Centaur Vitamin D Assay (500 tests); Reference Number:10631021. For use in the quantitative determination of Vit D.
Siemens Healthcare Diagnostics
Negative bias in patient samples for Vitamin D
The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.
When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.