πŸ”¬

Diagnostic Equipment

πŸ₯ Medical Devices β€’ 6,554 recalls

Siemens Healthcare Diagnostics has initiated a voluntary field action for Dimension(R) PHOS Flex(R) Reagent due to reagent interaction causing falsely depressed Creatinine results when EZCR is processed from open wells that are in close proximity to open wells of PHOS reagent.

Feb 22, 2013 Diagnostic Equipment Nationwide View Details β†’

Beckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley rubbing against the light shield can fall into the reaction vessels (RVs) and present a risk of erroneous test results.

Feb 12, 2013 Diagnostic Equipment Nationwide View Details β†’

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Feb 18, 2013 Diagnostic Equipment Nationwide View Details β†’

Medtronic has identified an issue involving specific lot numbers of the Medtronic Archer 0.035 inch (0.89mm) Super Stiff Guidewire, Models ARCSJ200W, ARCSJ260W, and ARCDJ260W, where a change in the manufacturing process has been identified as the cause for potential kinking and breaking at the proximal weld end of the outer spring coil during in-vivo use.

Mar 13, 2013 Diagnostic Equipment Nationwide View Details β†’

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Feb 18, 2013 Diagnostic Equipment Nationwide View Details β†’

Cartridges are not sealing properly during PCR testing and may cause false results, indeterminates and/or invalid runs.

Dec 11, 2012 Diagnostic Equipment View Details β†’

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Feb 18, 2013 Diagnostic Equipment Nationwide View Details β†’

Vital Images has found a potential error in measurement generated through the software which affects all versions of Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3. This error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/axia

Feb 5, 2013 Diagnostic Equipment View Details β†’

Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.

Nov 8, 2012 Diagnostic Equipment Nationwide View Details β†’

Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values

Nov 27, 2012 Diagnostic Equipment View Details β†’

Risk that air may be introduced into the device which may lead to an air embolism.

Feb 27, 2013 Diagnostic Equipment Nationwide View Details β†’

A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.

Jan 21, 2013 Diagnostic Equipment Nationwide View Details β†’

Natus Neurology Incorporated is recalling the Nicolet EEG Wireless Amplifier which may overheat and become uncomfortable to touch around the area of the on/off button. The overheating may result in discomfort ranging from too hot up to the potential for a burn.

Feb 27, 2013 Diagnostic Equipment Nationwide View Details β†’

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Dec 13, 2012 Diagnostic Equipment View Details β†’

Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.

Jan 28, 2013 Diagnostic Equipment View Details β†’