Diagnostic Equipment

🏥 Medical Devices 6,554 recalls

Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.

Oct 16, 2015 Diagnostic Equipment Nationwide View Details →

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

Sep 16, 2015 Diagnostic Equipment Nationwide View Details →

It would be possible for an abnormal analytical test result value from an affected newborn to be reported with a default result interpretation of normal. The analytical test result value would be correct but the interpretation of the test result value would be incorrect.

Aug 24, 2015 Diagnostic Equipment View Details →

May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.

Sep 16, 2015 Diagnostic Equipment Nationwide View Details →

ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the respective electrode is left unplugged after replacement, washing or maintenance activities. The calibration and 2-level QC will fail if the electrode is left unplugged. The IFU states to calibrate the ISEs after replacement and to run two levels of controls before running patient samples.

Aug 28, 2015 Diagnostic Equipment View Details →

Siemens will provide all customers a new version of system software (VA35E or VB10C depending on the hardware configuration). Hardware fixes will be implemented to address situations where the metal component used to engage the wheel lock becomes disconnected, and where the system locks up due to misalignment between the bezel and MPI board, in all units that have not already been corrected.

Oct 7, 2015 Diagnostic Equipment View Details →

In spine clinical workflows, cross reference lines may be used to determine the position of slices. In cases, where MobiView fused Images are used to show the cross reference lines, the cross reference lines may be positioned incorrectly.

May 5, 2014 Diagnostic Equipment Nationwide View Details →

When using the NicoletOne v5.94 software, after exiting the impedance check function and returning to the EEG screen, the impedance check signal remains active in waveform, obscuring the EEG signals.

Sep 11, 2015 Diagnostic Equipment Nationwide View Details →

Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.

Aug 27, 2015 Diagnostic Equipment Nationwide View Details →

This update is intended to provide a software update for the Syngo RT Oncologist, which is currently running SW version 4.2 or 4.3. The safety-related issues, which were described in the Customer Safety Notice distributed as UFSN-RTT/RTO 4.3 Adaptive Targeting [Auto-registration]. The update contains several important safety and performance fixes.

Aug 11, 2015 Diagnostic Equipment View Details →

Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.

Sep 10, 2015 Diagnostic Equipment Nationwide View Details →

RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results

Aug 24, 2015 Diagnostic Equipment Nationwide View Details →

RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results

Aug 24, 2015 Diagnostic Equipment Nationwide View Details →

RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results

Aug 24, 2015 Diagnostic Equipment Nationwide View Details →