DHEA-S causes falsely elevated progesterone results.
Diagnostic Equipment
๐ฅ Medical Devices โข 6,554 recalls
DHEA-S causes falsely elevated progesterone results.
The software is not adding the correct tomo image.
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
The probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
BD AffirM VPIII Microbial Identification Test
Becton Dickinson &
BD has confirmed that a portion of tests associated with the affected lots( 446252 and 446257 ) of product may exhibit blue speckles or fibers on the beads and cards, as shown below. This issue would result in an invalid test, as described in the package insert.
The monitor may not alarm appropriately for a pediatric or neonatal patient.
Potential performance issues. False Susceptible result instead of Intermediate and False Intermediate result with certain methods.
CSI discovered that a 1.50mm Solid Crown OAD was labeled as a 1.50mm Crown OED.
AMT 12 inch Right Angle Feeding Set with Y-Port Used with Nasogastric/Nasointestinal (NG/NI) tubes.
Applied Medical Technology
Expired expiration dates are listed on the pouches.
Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.
Tosoh Bioscience
The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.
Quidel Corporation is recalling Influenza Transport Swab Packages because it may include an incorrect sample swab mixed in with the proper swabs.
AMT Bridle - Nasal Tube Retaining System Used with Nasogastric/Nasointestinal (NG/NI) tubes.
Applied Medical Technology
Expired expiration dates are listed on the pouches.
The firm confirmed a reduced once-opened and on-board stability for N Latex Homocysteine (HCY) OPAX03 lot 802907632 that may result in erroneously reduced or elevated homocysteine values. The stability issue may lead to higher than expected lot-to-lot variation and an impaired product performance.
The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimensionยฟ clinical chemistry system. For prescription use only.
Siemens Healthcare Diagnostics
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Siemens Healthcare Diagnostics
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.