Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Diagnostic Equipment
π₯ Medical Devices β’ 6,554 recalls
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Siemens Healthcare Diagnostics
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8ΒΏC, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
AXIOM Sensis, Computer, Diagnostic, Programmable. Intended to be used as a diagnostic and administrative tool supporting hemodynamic cardiac catheterizations and/or intracardiac electrophysiology studies.
Siemens Medical Solutions USA
An extremely dusty computer can cause problems at system start, or rarely, cause system fail. Perform a system check prior to performing exams. If the Sensis system fails, its functions can not be used. Clinical treatment may need to be terminated, restarted, or transferred to a functioning system.
Potential performance issue on strain categorization.
Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010
Magellan Diagnostics
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Access BR Monitor, Part No. 387620
Beckman Coulter
Following biases observed in the results of the French National Quality Controls, Beckman Coulter has been informed that a study completed by the French Competent Authority Agence nationale de sΒΏcuritΒΏ du Medicament et des produits de santΒΏ (ANSM) showed diagnosed cancer patients in non-remission of their metastatic disease, may have results below the upper reference limit (URL) of 31.3 U/mL as stated in the Access BR Monitor instructions for use (IFU).
Ziehm Imaging solo FD Mobile Fluoroscopic C-arm
Ziehm Imaging
During Ziehm Imaging GmbH FDA inspection, the label for the Collimator and X-ray Control panel was missing the manufacturer date as part of the identification for certified component
ADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the quantitative determination of insulin in serum.
Siemens Healthcare Diagnostics
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010
Magellan Diagnostics
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
False susceptible results
AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.
Abbott Medical Optics Inc. (AMO)
Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.
ADVIA Centaur Systems Insulin Calibrator; Test Code: Calibrator IRI; Catalog Number: 04618899
Siemens Healthcare Diagnostics
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.
Hitachi Medical Systems America
The strength of the arm mounting mechanism was insufficient and had to be redesigned.
Potential Patient Demographic Error with Blank Patient ID Field