A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Diagnostic Equipment
🏥 Medical Devices • 6,554 recalls
The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.
The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.
The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 867146
Philips Electronics North America
Five warning statements are missing from the instructions for use.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
IntelliVue MX40 Patient Monitor, Software Versions B.05, B.06, B.06.5X, Model 865351
Philips Electronics North America
Five warning statements are missing from the instructions for use.
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Users have misinterpreted the display for out of range measurement indicated by the blinking" ---" to mean a zero measurement.
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.
RX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical chemistry tests.
Randox Laboratories
Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results.
The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Users have misinterpreted the display for out of range measurement indicated by the blinking "-0-" to mean a zero measurement
While in Full Disclosure playback, a user may inadvertently close the Full Disclosure Control Window using the ESC key function, instead of pressing the X icon in the upper right corner of this window.
Report was of an interrupted MR scan after patient complained about acoustic distortion, and heat felt form the headphones.
Replacement of units lacking an updated device approval.