The device and its accessories may not have been manufactured according to current Good Manufacturing Practices. In addition, the firm has received three complaints regarding electric shocks received when using the device.
Diagnostic Equipment
🏥 Medical Devices • 6,554 recalls
Potential polymer material degradation.
INOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020).
There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
The software is not identifying the patient as having atypical hyperplasia, resulting in an incorrect Gail Risk calculation.
There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
Diamedix Is-Mumps IgG Enzyme Immunoassay Test Kit for In Vitro Diagnostic Use Catalog No. 720-540
Diamedix
Contains a gel-like contaminant that may affect product performance.
Diamedix Is-EBV-EA-D IgG Enzyme Immunoassay Test Kit for In Vitro Diagnostic Use Catalog No. 720-640
Diamedix
Contains a gel-like contaminant that may affect product performance.
Underfilled reagent well that could lead to incorrect results
There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.
The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.
There is a potential for the knob assembly in the swivel adapter to fracture/break during use.
An improper wiring connection on the chiller assembly may result in overheating of the connection and potential failure of certain components in the system.
The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
The placing of weight on the VRAD detector, of the Brivo XR385 digital radiographic system, can introduce image artifacts, which can result in patient exam retakes.
AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.
Siemens Medical Solutions USA
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.
Siemens Medical Solutions USA
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters affect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must be aware to avoid incorrect dose calculations during treatment planning.