Adding an additional comparison study to the Clinical Side Bar in the Enterprise Imaging Diagnostics Desktop, can result in a change of the sort order of the studies in the Clinical Side Bar. The thumbnail images of the studies are no longer chronologically sorted with the most recent study at the top.
Diagnostic Equipment
🏥 Medical Devices • 6,554 recalls
The device may not function at the lower end (<15 deg. C) of the labeled operating temperatures. The instrument returns to normal operating conditions when environmental temperatures are at or above 15 deg. C. This issue has no potential to cause harm to either the patient or the operator. At the lower temperature, the instrument may fail to initialize, in which case slides would require manual screening. Therefore, no slides are screened by the device and no results can be released in error. The instrument labeling has been revised to indicate an operating temperature lower limit of 18 deg. C and operating humidity of 20 to 80% relative humidity.
VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266
Ortho Clinical Diagnostics
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
"Manufactured" and "Use By" labels are transposed incorrectly suggesting the product expires before it was manufactured.
VITROS Immunodiagnostic Products - FSH Reagent Pack (10758750000302) 1931922
Ortho Clinical Diagnostics
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
VITROS Immunodiagnostic Products - Prolactin Reagent Pack (10758750000111) 1849793
Ortho Clinical Diagnostics
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
RTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCB
VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997
Ortho Clinical Diagnostics
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.
Leica Microsystems
Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.
"Manufactured" and "Use By" labels are transposed incorrectly suggesting the product expires before it was manufactured.
AIIuraXperFDlO/10, System Code 722011 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
INTEGRIS CV, System Code 722030 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
Allura Xper FD1O C, System Code 722001 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
Poly C- DMCP-Visub(H3000), System Code 72238 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
INTEGRIS SUITE, System Code 722199 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
AIIura Xper F010 DR Table, System Code 722022 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
Allura Xper FD2O, System Code 722006 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
INTEGRIS Allura 9 (biplane), System Code 722021 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
Allura Xper FD2O, System Code 722012 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
INTEGRIS Allura 15-12 (mono), System Code 722043 Vascular, cardiovascular and neurovascular imaging applications, Cardiac imaging applications including diagnostics, Non-vascular interventions
Philips Electronics North America
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.