When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Diagnostic Equipment
π₯ Medical Devices β’ 6,554 recalls
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System.
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin in human serum and plasma on Roche/Hitachi cobas c systems.
Roche Diagnostics
The reagent lot is exhibiting calibration failures, QC imprecision, and elevated QC recovery due to elevated pH level causing higher absorbance values.
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
Edwards Lifesciences has received a limited number of customer reports regarding an inter-lumen leak between the cardioplegia lumen and pressure lumen in the device.
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.