When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Mar 6, 2019 Diagnostic Equipment Nationwide View Details β†’

There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.

Mar 15, 2019 Diagnostic Equipment Nationwide View Details β†’

Foot Switches used with the following systems: 722001 Allura Xper FD 10 C 722002 Allura Xper FD10 F 722003 Allura Xper FD10 722005 Allura Xper FD10/10 722006 Allura Xper FD20 722008 Allura Xper FD20 Biplane 722010 Allura Xper FD10 722011 Allura Xper FD10/10 722012 Allura Xper FD20 722013 Allura XPER FD20 BIPLANE 722014 Allura Xper FD10 OR Table 722015 Allura Xper FD20 OR Table 722019 Allura Xper FD10/10 OR Table 722020 Allura Xper FD20 Biplane OR Table 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722024 Allura Xper FD10/10 OR Table 722025 Allura Xper FD20 Biplane OR Table 722026 Allura Xper FD10 722027 Allura Xper FD10/10 722028 Allura Xper FD20 722029 Allura Xper FD20/10 722031 Allura CV20 722033 Allura Xper FD10 OR Table 722034 Allura Xper FD10/10 OR Table 722035 Allura Xper FD20 OR Table 722036 Allura Xper FD20/10 OR Table 722038 Allura Xper FD20/20 722039 Allura Xper FD20/20 OR Table 722058 Allura Xper FD20/15 722059 Allura Xper FD20/15 OR Table 722123 Field extensions Xper cardio 722124 Field extensions Xper vascular 722126 SmartPath to upgrade Allura to AlluraClarity cardio 722127 SmartPath to upgrade Allura to AlluraClarity vascular 722133 Field ext. Xper cardio systems R7.6 722134 Field ext. Xper vascular systems 722400 Cardio Vascular-Allura Centron 722030 INTEGRIS CV 722043 INTEGRIS Allura 15-12 (mono) Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

Philips Medical Systems Nederlands

Class I - Dangerous

Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

Feb 19, 2019 Diagnostic Equipment Nationwide View Details β†’

The safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System.

Feb 26, 2019 Diagnostic Equipment Nationwide View Details β†’

There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.

Mar 15, 2019 Diagnostic Equipment Nationwide View Details β†’

New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***

Dec 17, 2018 Diagnostic Equipment Nationwide View Details β†’

There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.

Mar 15, 2019 Diagnostic Equipment Nationwide View Details β†’

Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.

Mar 12, 2019 Diagnostic Equipment Nationwide View Details β†’

COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over

Dec 14, 2018 Diagnostic Equipment Nationwide View Details β†’

There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.

Jul 17, 2018 Diagnostic Equipment Nationwide View Details β†’

COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over

Dec 14, 2018 Diagnostic Equipment Nationwide View Details β†’

There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.

Jul 17, 2018 Diagnostic Equipment Nationwide View Details β†’

The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.

Feb 7, 2019 Diagnostic Equipment View Details β†’

There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.

Jul 17, 2018 Diagnostic Equipment Nationwide View Details β†’

Edwards Lifesciences has received a limited number of customer reports regarding an inter-lumen leak between the cardioplegia lumen and pressure lumen in the device.

Jan 16, 2019 Diagnostic Equipment Nationwide View Details β†’

There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.

Mar 21, 2019 Diagnostic Equipment Nationwide View Details β†’

There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.

Mar 21, 2019 Diagnostic Equipment Nationwide View Details β†’