The Anti-Fog Solution may not defog properly due to a degradation in the chemical properties.
Diagnostic Equipment
π₯ Medical Devices β’ 6,554 recalls
Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4D
GE Healthcare
Potential for a transducer mis-alignment in certain transvaginal probes.
Additional low dose radiation exposure.
The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.
The products have been confirmed to have reduced or no additive within the tube reservoir. The tubes may develop visible clots within the tube samples or micro clots that are not easily detected during visual inspection of the tubes. A clotted sample may lead to recollection of samples or, retesting of patients, resulting in delayed reporting of test results and patient treatment.
software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.
ColoCARE, Cat. No. 5651, packaged 50 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
Helena Laboratories
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
ColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
Helena Laboratories
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
A steady upward trend in blank (u) absorbance was observed
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
There is a potential for the device to fail to deliver energy to the patient.
Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.
Roche Diagnostics
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
A steady upward trend in blank (u) absorbance was observed