The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
Zeus Scientific
The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
Assay files used on the open Dynex DSX Instrument to process IVD ELISA samples-programming error results in well H6 being skipped for the addition of the TMB substrate. Well H6 will generate a negative result regardless of whether the patient was negative or positive for that antibody. This could result in a potential false negative result for that assay for any patients tested in well H6.