Zeus Scientific

4 recalls Categories: Diagnostic Equipment, Infusion Pumps

FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.

Aug 12, 2019 Diagnostic Equipment Nationwide View Details →

Assay files used on the open Dynex DSX Instrument to process IVD ELISA samples-programming error results in well H6 being skipped for the addition of the TMB substrate. Well H6 will generate a negative result regardless of whether the patient was negative or positive for that antibody. This could result in a potential false negative result for that assay for any patients tested in well H6.

Jul 22, 2019 Infusion Pumps Nationwide View Details →