Zap Surgical Systems

2 recalls • • Categories: Other Medical Devices

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Jul 25, 2025 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

Jul 5, 2022 Other Medical Devices Nationwide View Details →