Viewray

6 recalls • • Categories: Infusion Pumps, Diagnostic Equipment, Other Medical Devices

Software anomalies affecting the French, German and Italian versions of treatment delivery system (TDS) software.

May 4, 2021 Infusion Pumps View Details →

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

Oct 20, 2020 Diagnostic Equipment Nationwide View Details →

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

Oct 20, 2020 Diagnostic Equipment Nationwide View Details →

Registering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy and structures. The MRIdian TPDS software shows this misalignment to the user in the predicted dose-volume histogram (DVH). The misalignment impact may not be obvious to the user. During plan re-optimization the previously delivered dose would be incorrectly accounted for by the software. This may result in unexpectedly higher or lower dose than the intended dose calculated for the subsequent treatment plan. When this occurs in the treatment planning workflow, the registration fails to align the dose with the image and the user is unable to proceed.

Nov 18, 2019 Infusion Pumps View Details →