Verathon

14 recalls Categories: Diagnostic Equipment, Infusion Pumps, Other Medical Devices
Class I - Dangerous

Handheld video monitor failed to meet the labeled IP67 rating which may allow fluid ingress into the monitor resulting in loss of functionality.

Jul 1, 2021 Diagnostic Equipment Nationwide View Details →

When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Jun 5, 2020 Diagnostic Equipment Nationwide View Details →

Product labeled as GVL 2 Stat (blades), lot GC32825, was physically packaged with the incorrect size device (GVL 1 Stat (blades)).

Mar 3, 2017 Other Medical Devices Nationwide View Details →

The firm is providing customers with an updated Operations and Maintenance Manual for each of the BladderScan BVI 9600 and AortaScan AMI 9700 devices. The updated Operations and Maintenance Manuals clarify that these devices should not be used for the screening, detection, or diagnosis of abdominal aortic aneurysms (AAAs).

Nov 30, 2015 Other Medical Devices Nationwide View Details →

The firm has become aware of the potential for disruption (flickering) in the video laryngoscopy image when GlideScope Titanium SU laryngoscope blades are in use. Video "flickering" appears as the intermittent break-up of the on-screen video image, appearing as either distorted horizontal or vertical bars of displaced video signal.

Jan 29, 2016 Diagnostic Equipment Nationwide View Details →

Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.

Apr 13, 2015 Diagnostic Equipment View Details →

Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine inspection before or after intubation.

Nov 30, 2015 Diagnostic Equipment Nationwide View Details →

Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine inspection before or after intubation.

Nov 30, 2015 Diagnostic Equipment Nationwide View Details →

Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.

May 8, 2013 Diagnostic Equipment Nationwide View Details →

Re-usable video laryngoscope blades of the GlideScope GVL3, GVL4, and GVL5 were recalled due to potential cracking.

Oct 25, 2012 Diagnostic Equipment Nationwide View Details →