Trilliant Surgical

23 recalls Categories: Implants & Prosthetics, Other Medical Devices, Surgical Instruments, Infusion Pumps

The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.

Nov 4, 2019 Other Medical Devices Nationwide View Details →

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018 Surgical Instruments Nationwide View Details →

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018 Surgical Instruments Nationwide View Details →

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018 Surgical Instruments Nationwide View Details →

Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.

Dec 8, 2016 Other Medical Devices Nationwide View Details →

Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.

Sep 2, 2016 Surgical Instruments View Details →

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Jul 31, 2015 Infusion Pumps View Details →

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Aug 6, 2015 Infusion Pumps View Details →

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Jul 20, 2015 Infusion Pumps Nationwide View Details →

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Jul 20, 2015 Implants & Prosthetics Nationwide View Details →

The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

Jul 20, 2015 Implants & Prosthetics Nationwide View Details →

The cutting efficiency of the Countersinks was out of specification.

Jul 20, 2015 Implants & Prosthetics Nationwide View Details →