Tosoh Smd

3 recalls • • Categories: Diagnostic Equipment, Infusion Pumps, Other Medical Devices

The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898.

Nov 15, 2019 Diagnostic Equipment Nationwide View Details →

Power outages causes reporting software to shutdown.

Dec 22, 2016 Infusion Pumps Nationwide View Details →

Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.

May 15, 2013 Other Medical Devices Nationwide View Details →