Spiegelberg Gmbh & Co. KG

1 recall • • Categories: Diagnostic Equipment

In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.

Dec 29, 2009 Diagnostic Equipment Nationwide View Details →