Skytron, Div. The KMW Group

9 recalls Categories: Other Medical Devices, Diagnostic Equipment, Surgical Instruments

Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.

Mar 10, 2020 Other Medical Devices Nationwide View Details →

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

Apr 6, 2016 Surgical Instruments Nationwide View Details →

The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket. A display interface falling from its mounting location may result in injury to the patient or provider.

Dec 2, 2015 Diagnostic Equipment Nationwide View Details →