One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.
SIGHT DIAGNOSTICS
Class I - Dangerous
Class I - Dangerous
The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software version 2.63, 2.63.1, and 2.63.1.1. Due to the issue, results may incorrectly appear to be within or outside the reference range.
In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on software versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing incorrectly-configured reference range values
Class I - Dangerous
Discrepancy in labeling between the adult CBC reference ranges presented in the OLO Operator's Manual and the OLO software