6 recalls Categories: Other Medical Devices, Diagnostic Equipment, Surgical Instruments

Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft.

Oct 16, 2025 Other Medical Devices Nationwide View Details →

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Sep 6, 2024 Other Medical Devices Nationwide View Details →

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Sep 6, 2024 Diagnostic Equipment Nationwide View Details →

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

Sep 6, 2024 Surgical Instruments Nationwide View Details →

Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.

Apr 10, 2023 Other Medical Devices Nationwide View Details →