Radiometer America

25 recalls Categories: Other Medical Devices, Diagnostic Equipment, Infusion Pumps

Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.

Jul 10, 2019 Other Medical Devices Nationwide View Details →

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.

Jul 2, 2018 Other Medical Devices Nationwide View Details →

The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again

May 25, 2018 Other Medical Devices View Details →

Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.

Oct 24, 2017 Diagnostic Equipment Nationwide View Details →

Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kPa or 36.9  46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

Sep 7, 2017 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Reports that the sample type "cord blood" has been changed into the two sample types "cord blood arterial" and "cord blood venous.

Jan 26, 2017 Other Medical Devices Nationwide View Details →

Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.

Nov 23, 2016 Diagnostic Equipment View Details →
Class I - Dangerous

Radiometer America, Inc. is recalling TCM CombiM on rare occasions they may have an error that can cause the alarm to sound higher than the pre-set maximum volume.

Nov 18, 2015 Other Medical Devices Nationwide View Details →

When the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Jul 16, 2015 Diagnostic Equipment View Details →

Negative drift on QC during in-use lifetime of CREA membranes can be observed.

Oct 29, 2014 Diagnostic Equipment Nationwide View Details →

Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag

Oct 20, 2014 Infusion Pumps Nationwide View Details →

RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.

Jun 10, 2013 Diagnostic Equipment Nationwide View Details →

A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t

Jul 5, 2012 Diagnostic Equipment Nationwide View Details →

A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t

Jul 5, 2012 Diagnostic Equipment Nationwide View Details →

A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t

Jul 5, 2012 Diagnostic Equipment Nationwide View Details →