Higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument
Phadia US
Higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument
Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).
The stability specifications of ImmunoCAP Allergen c5, Ampicilloyl were not met.
The stability specifications of ImmunoCAP Allergen c6, Amoxicilloyl were not met.
The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.
Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles. Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.
Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step, when combined with a change in material source in kit lot #0067, caused elevated results.