Phadia Ab

5 recalls Categories: Diagnostic Equipment, Surgical Instruments

Potential for reporting low assay results

May 10, 2019 Surgical Instruments Nationwide View Details →

We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU).

Nov 20, 2017 Surgical Instruments View Details →

All equivocal and positive results (greater or equal 7 EliA U/ml) on EliA SmDP well lot 0018 may be incorrect and must be considered invalid. Patient samples in complaint investigations caused unspecific signals up to 22 U/ml on EliA SmDP well lot 0018. The nonspecific signals are not caused by anti-Sm antibodies, nor by streptavidin antibodies.

Dec 28, 2015 Surgical Instruments Nationwide View Details →