PerkinElmer 226 Sample Collection Devices Lot Number 40020001 had serial numbers printed on the pack labels that did not match the serial number range of the devices contained within the pack.
PerkinElmer Health Sciences
PerkinElmer 226 Sample Collection Devices Lot Number 317114 had serial numbers printed on the pack labels that did not match the serial number range of the devices contained within the pack.
AutoDELFIA System with Version 3.0 Software: Potential plate loading failures
AutoDELFIA System with Version 3.0 Software: Potential plate loading failures
AutoDELFIA System with Version 3.0 Software: Potential plate loading failures
Issue may cause eReports to consume an excessive amount of disk space causing the software to stop operating.
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The change observed in the serum samples within the range 1.76 67.0 IU/L hFSH averaged -2.9%. The reference range of the assay spans 0.60 to 98.0 IU/L.
Unique Device Identifier (UDI) is missing from outer kit box label.
Potential for demographic information and test result to be mismatch.
Risk of incorrect results from samples prepared with Panthera-Puncher 9