NRT X-RAY A/S

4 recalls Categories: Other Medical Devices, Infusion Pumps

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.

Apr 17, 2024 Other Medical Devices Nationwide View Details →