A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.
Nova Biomedical
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.
Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Reports of false abnormally high glucose readings from test strips
Glucose test strips reports no result message when tested
Glucose test strips reports no result message when tested