Mizuho OSI

6 recalls Categories: Other Medical Devices, Surgical Instruments

Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.

Mar 28, 2017 Other Medical Devices View Details →

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.

Dec 14, 2015 Other Medical Devices View Details →