MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK

6 recalls Categories: Other Medical Devices

In rare cases the VERO/MHI-TM2000 Operator Console could set an incorrect Gantry angle for the first beam of an IMRT treatment if certain specific conditions are met. If Gantry angle for the first beam of IMRT is incorrectly set and if nobody is aware of it, an erroneous treatment would be given to the patient.

Mar 7, 2016 Other Medical Devices Nationwide View Details →

Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion. As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port

Dec 2, 2014 Other Medical Devices View Details →

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

Jul 5, 2014 Other Medical Devices View Details →

Software anomaly: Mitsubishi Heavy Industries, Ltd. became aware of a potential safety issue associated with the VERO (TM) Linear Accelerator System. The VERO (TM) may save treatment delivery record into wrong location under certain conditions.

Jun 21, 2013 Other Medical Devices Nationwide View Details →