Mevion Medical Systems

14 recalls Categories: Other Medical Devices, Infusion Pumps
Class I - Dangerous

Treatment is allowed to continue (via partials) in situations where dDose1 interlock has tripped

Oct 1, 2018 Other Medical Devices View Details →

QFix has advised Mevion Medical Systems that they are identifying unauthorized combinations of QFix Couch Tops with non-QFix devices in clinical use. The kVue Oen Touch latch consists of a precision latch mechanism that can be damaged when used with incompatible (third-party) accessories. This can create unsafe conditions for use and may compromise the subsequent use of any kVue inserts, whether authorized or not. Use of unauthorized or unvalidated inserts may lead to an incorrect patient treatment or physical injury.

Jun 29, 2018 Infusion Pumps Nationwide View Details →

Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.

Mar 15, 2016 Other Medical Devices Nationwide View Details →

The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.

Nov 18, 2015 Other Medical Devices Nationwide View Details →

Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%.

Nov 10, 2014 Other Medical Devices View Details →