Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
Medica
3 recalls • • Categories: Diagnostic Equipment, Other Medical Devices, Infusion Pumps
Class I - Dangerous
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.