The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.
Mallinckrodt
6 recalls • • Categories: Infusion Pumps, Prescription Drugs, Surgical Instruments
🏥 Medical Devices Class I - Dangerous
Labeling: incorrect or missing lot number and/or expiration date
Labeling: Correct Labeled Product Miscart/Mispack: labels on outer containers do not match labels on vials (the correct label)
💊 Drugs Class I - Dangerous
CGMP Deviations: This product is being recalled because expired flavoring was used in the manufacturing of these lots.
💊 Drugs Class I - Dangerous
Subpotent (Multiple Ingredient) Drug: Low out of specification assay results for the hydrocodone bitartrate ingredient was found.
🏥 Medical Devices Class I - Dangerous
A report was received from a Japanese distributor regarding a complaint from a hospital customer stating that ten packages of Handi Fil were not sealed.