Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than ¿ 35 percent. The dosage displayed on the generator console and table display (monitor) is higher than the actual delivered dose.
LIEBEL-FLARSHEIM COMPANY
Class I - Dangerous
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.