Potential that the induction seal is not completely sealed to the device tube packaging.

Jan 10, 2025 Infusion Pumps Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods

Sep 20, 2024 Surgical Instruments Nationwide View Details →

Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.

Jul 9, 2024 Surgical Instruments Nationwide View Details →

Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal that may compromise the product sterility.

May 7, 2024 Surgical Instruments Nationwide View Details →

Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.

May 7, 2024 Surgical Instruments Nationwide View Details →

Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.

May 7, 2024 Surgical Instruments Nationwide View Details →

Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.

Oct 5, 2023 Diagnostic Equipment Nationwide View Details →

Released with discoloration (rust) on the Drill Chuck and inner chuck of Codman Cranial Hand Drill may result in inflammation, local toxicity, systemic toxicity, sensitivity, revision surgery, procedural delay, inconvenience to user, and/or infection.

Apr 11, 2023 Diagnostic Equipment View Details →

lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.

Aug 4, 2020 Surgical Instruments Nationwide View Details →

lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.

Aug 4, 2020 Surgical Instruments Nationwide View Details →

lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.

Aug 4, 2020 Surgical Instruments Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Oct 21, 2019 Other Medical Devices Nationwide View Details →