Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
Imactis
Class I - Dangerous
Class I - Dangerous
The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.
The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.