Horiba Instruments Incorporated

2 recalls Categories: Surgical Instruments, Other Medical Devices

HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.

Mar 28, 2024 Surgical Instruments Nationwide View Details →

Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).

Oct 30, 2018 Other Medical Devices Nationwide View Details →