Getinge Group Logistics America

4 recalls • • Categories: Other Medical Devices, Surgical Instruments, Patient Monitors
Class I - Dangerous

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Feb 11, 2021 Other Medical Devices Nationwide View Details →

Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.

Jan 29, 2021 Other Medical Devices Nationwide View Details →

The potential that the plastic inlet hose connections may deform or break as a result of material fatigue and/or over tightening of the clamps. A broken hose connection may lead to a hazardous situation of water leakage or water spray. The water leakage or water spray may result in a burn to the user and/or an inconvenience due to a procedural delay if the Washer Disinfector is taken out of service.

Dec 21, 2020 Surgical Instruments Nationwide View Details →

A potentially shorter than specified nebulizer connector may result in difficulty in installing the Aeroneb Nebulizer module to the Servo-i device successfully.

Sep 17, 2020 Patient Monitors View Details →