Getinge Disinfection Ab

7 recalls Categories: Other Medical Devices, Surgical Instruments

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Apr 2, 2025 Other Medical Devices Nationwide View Details →

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Apr 2, 2025 Other Medical Devices Nationwide View Details →

There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

Mar 2, 2018 Other Medical Devices Nationwide View Details →

There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.

Mar 2, 2018 Other Medical Devices Nationwide View Details →

Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.

Nov 26, 2014 Surgical Instruments Nationwide View Details →