Software versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence
Fujifilm Medical Systems U.S.A.
Potentially defective Chip Cassette units, when used for DCP or AFP-L3% measurement, incorrect DCP and AFP-L3% values may result
The wrong patient information may be displayed in the viewer or PowerJacket.
FUJIFILM has become aware that incorrect concentration values were printed on the bottle label for all product manufactured as part of Lot EL629. This may result incorrect results when used in testing.
The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.
There is a potential for the wrong patient information may be displayed in the viewer or PowerJacket. Additionally, the incorrect max standard uptake values (SUV) for PET ad CT studies may be returned for Siemens Modality.
FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected.
Lateromedial (LM) and Lateromedial Oblique (LMO) Orientation Markers may be displayed incorrectly on the 3D TOMO slider bar. In addition, the slice location of the image and/or the slice direction could be incorrect as well, and result in misdiagnosis
The graphics driver of the FDR Go PLUS might cause the appearance of a Blue Screen of Death (BSoD)
FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
There was a case that the swivel arm went down unintentionally when it was moved up/down.
There was a case that the swivel arm went down unintentionally when it was moved up/down.
A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device. The action includes replacement of the forceps elevator mechanism, the O-ring seal, the distal end cap and issuance of a new Operation Manual.
Fujifilm is conducting a corrective action due to an FDA inspection.
Fujifilm is reporting a corrective action due to an FDA inspection.
Fujifilm is recalling multiple endoscopes after a retrospective review.
Image data for a patients image may not be correct
If an exposure is aborted during a stereo biopsy workflow, an error may occur which would prevent the system from continuing to operate properly.
Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures.
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures.
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures.
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures.
This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures.
Reports of failure of the device's power supply.
A precautionary measure because the Duodenoscopes have been redesigned to incorporate a closed elevator wire channel that facilitates reprocessing.
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.
The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures.
Under certain circumstances the patient Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report.
Measurement values in the EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report.
User Manual lacks a description for handling some errors even though the system can be recovered from such errors