Flowonix Medical

13 recalls Categories: Infusion Pumps, Implants & Prosthetics

Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection

May 26, 2021 Infusion Pumps Nationwide View Details →

A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.

Dec 24, 2019 Implants & Prosthetics Nationwide View Details →

A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.

Dec 24, 2019 Implants & Prosthetics Nationwide View Details →

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.

Mar 21, 2019 Infusion Pumps Nationwide View Details →

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.

Mar 21, 2019 Infusion Pumps Nationwide View Details →

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.

Mar 21, 2019 Infusion Pumps Nationwide View Details →

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.

Mar 21, 2019 Infusion Pumps Nationwide View Details →

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.

Mar 21, 2019 Infusion Pumps Nationwide View Details →

Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to permit safe exposure to an MRI without removing drug from the reservoir.

May 22, 2017 Implants & Prosthetics Nationwide View Details →