Extremity Medical

4 recalls • • Categories: Surgical Instruments, Other Medical Devices, Implants & Prosthetics

The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.

Jan 10, 2013 Surgical Instruments View Details →

Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).

Mar 22, 2012 Other Medical Devices Nationwide View Details →

Product Event Reports revealed that lag screw passes through washer (e.g. X-Post) intra-operatively.

Apr 13, 2011 Implants & Prosthetics Nationwide View Details →