Cytocell

14 recalls • • Categories: Other Medical Devices, Diagnostic Equipment

Device is mislabeled. The package insert for this ASR device contains a chromomap - an image showing the position the FISH probe maps to on the specific chromosome. For this device the chromomap in version 1 of the package insert is incorrect and doesn't not match the probe design or specification contained on the package insert

Oct 27, 2022 Other Medical Devices View Details →

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Jan 18, 2022 Diagnostic Equipment Nationwide View Details →

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Dec 22, 2020 Other Medical Devices View Details →

A low risk of a false positive result being issued with a laboratory developed test (LDT) that utilizes the LPH533-A NUP98 Distal Probe Green.

Jul 17, 2020 Other Medical Devices Nationwide View Details →