DNAs used during the manufacture of two affected models were accidentally inverted during manufacturing.
Cytocell
DNAs used during the manufacture of two affected models were accidentally inverted during manufacturing.
An error with the chromomap on version 1 of the package insert which is incorrect. The nucleotide locations specified in the insert are correct but the map has been created using the wrong coordinates.
Device is mislabeled. The package insert for this ASR device contains a chromomap - an image showing the position the FISH probe maps to on the specific chromosome. For this device the chromomap in version 1 of the package insert is incorrect and doesn't not match the probe design or specification contained on the package insert
Labelled with an incorrect expiry date on the outer label with an extended expiry date than the validated shelf life. The correct expiry date for the device is 2022-02 and not 2023-05 as shown on the outer label.
May show unexpected locus specific signals in addition to those at 14q32.
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Cytocell LPH019-A E2A (TCF3) Breakapart Probe- In Vitro Diagnostic Analyte specific reagents (ASR's)
May show unexpected locus specific signals in addition to those at 19p13.3. Users may observe faint additional locus specific signals at 9q34
The device may show unexpected locus specific signals in addition to those at 22q. Users may observe faint additional locus specific signals at 20p12/13.
Users may observe additional locus specific signals at 9q34. The device pack insert indicates the probe has no known cross-reactivity to 9q34.
Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.
A low risk of a false positive result being issued with a laboratory developed test (LDT) that utilizes the LPH533-A NUP98 Distal Probe Green.
There is an error in the chromomap on the package insert.
There is an error in the chromomap included in the labeling for the product.